DL for lung cavity estimation from Xe- and H-MRI

A Dual-Channel Deep Learning Approach for Lung Cavity Estimation From Hyperpolarized Gas and Proton MRI

by Joshua R. Astley BEng, Alberto M. Biancardi PhD, Helen Marshall PhD, Paul J. C. Hughes PhD, Guilhem J. Collier PhD, Laurie J. Smith PhD, James A. Eaden PhD, Rod Hughes MD, Jim M. Wild PhD, Bilal A. Tahir PhD


JMRI (2022). doi: 10.1002/jmri.28519

Abstract

Background
Hyperpolarized gas MRI can quantify regional lung ventilation via biomarkers, including the ventilation defect percentage (VDP). VDP is computed from segmentations derived from spatially co-registered functional hyperpolarized gas and structural proton (1H)-MRI. Although acquired at similar lung inflation levels, they are frequently misaligned, requiring a lung cavity estimation (LCE). Recently, single-channel, mono-modal deep learning (DL)-based methods have shown promise for pulmonary image segmentation problems. Multichannel, multimodal approaches may outperform single-channel alternatives.

Purpose
We hypothesized that a DL-based dual-channel approach, leveraging both 1H-MRI and Xenon-129-MRI (129Xe-MRI), can generate LCEs more accurately than single-channel alternatives.

Study Type
Retrospective.

Population
A total of 480 corresponding 1H-MRI and 129Xe-MRI scans from 26 healthy participants (median age [range]: 11 [8–71]; 50% females) and 289 patients with pulmonary pathologies (median age [range]: 47 [6–83]; 51% females) were split into training (422 scans [88%]; 257 participants [82%]) and testing (58 scans [12%]; 58 participants [18%]) sets.

Field Strength/Sequence
1.5-T, three-dimensional (3D) spoiled gradient-recalled 1H-MRI and 3D steady-state free-precession 129Xe-MRI.

Assessment
We developed a multimodal DL approach, integrating 129Xe-MRI and 1H-MRI, in a dual-channel convolutional neural network. We compared this approach to single-channel alternatives using manually edited LCEs as a benchmark. We further assessed a fully automatic DL-based framework to calculate VDPs and compared it to manually generated VDPs.

Statistical Tests
Friedman tests with post hoc Bonferroni correction for multiple comparisons compared single-channel and dual-channel DL approaches using Dice similarity coefficient (DSC), average boundary Hausdorff distance (average HD), and relative error (XOR) metrics. Bland–Altman analysis and paired t-tests compared manual and DL-generated VDPs. A P value < 0.05 was considered statistically significant.

Results
The dual-channel approach significantly outperformed single-channel approaches, achieving a median (range) DSC, average HD, and XOR of 0.967 (0.867–0.978), 1.68 mm (37.0–0.778), and 0.066 (0.246–0.045), respectively. DL-generated VDPs were statistically indistinguishable from manually generated VDPs (P = 0.710).

Data Conclusion
Our dual-channel approach generated LCEs, which could be integrated with ventilated lung segmentations to produce biomarkers such as the VDP without manual intervention.